RecallDepth

Justright Surgical, JustRight 5mm Reload, REFJR-REL25-2.0, Sterile, EO, Rx Only It is a replacement cartridge used when more than one firing is required during a surgery.

This recall has been terminated (originally issued September 23, 2014).

Company
Justright Surgical, Llc
Recall Initiated
August 27, 2014
Posted
September 23, 2014
Terminated
October 6, 2014
Recall Number
Z-2691-2014
Quantity
3 cases containing 36 units
Firm Location
Boulder, CO
Official Source
View on FDA website ↗

Reason for Recall

JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.

Distribution

Distributed in the states of Colorado, California and New York.

Lot / Code Info

Lot: 75FD0417

Root Cause

Process design

Action Taken

Consignees were notified via letter on August 27, 2014 in addition, the sales rep will hand deliver the Recall Notification on 8/28/14. Acknowledgement Form will be provided and returned products will be sent to STD Med where it will be held in quarantine.