CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

This recall has been terminated (originally issued February 7, 2015).

Company
Maquet Datascope Corp Cardiac Assist Division
Recall Initiated
December 16, 2014
Posted
February 7, 2015
Terminated
June 9, 2017
Recall Number
Z-1062-2015
Quantity
838 units-US; 364 units - OUS
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.

Distribution

Worldwide Distribution

Lot / Code Info

Sensation 7 fr.34 cc IAB Demonstration Unit  0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469  Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470  Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units  0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568  Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576  Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272

Root Cause

Device Design

Action Taken

Maquet Inc. sent a Product Recall Medical Device Field Correction letter/Field Correction response form, dated 12/16/2014 via Fed Ex. The letter identified the affected product and the reason for the recall. Users were asked to inspect their inventory for the affected units. If customers have any affected product, they are to follow the instructions provided. Additionally, users are to complete and return the attached Field Correction Response Form to acknowledge receipt of the letter and to indicate completion of the steps outlined. Completed forms should be returned by e-mailing a scanned copy to fieldactions@maquet.com or faxing a copy to 1-973-396-3607. Technical questions should be directed to the Technical Support Department at 1-800-777-4222.