Cardiosave Rescue IABP, Part Number 0998-UC-0800-83

This recall is currently active, issued October 27, 2018. It was issued by Maquet Datascope Corp Cardiac Assist Division.

Company
Maquet Datascope Corp Cardiac Assist Division
Recall Initiated
September 20, 2018
Posted
October 27, 2018
Recall Number
Z-0100-2019
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.

Distribution

US Nationwide; Internationally to 77 countries

Lot / Code Info

UDI - 10607567108407 All Serial Numbers

Root Cause

Device Design

Action Taken

On September 20, 2018, Maquet/Getinge issued a press release. On September 24, 2018 and on September 28, 2018, Maquet/Getinge issued URGENT MEDICAL DEVICE CORRECTION notices along with an Instructions for Use Addendum to customers. On October 23, 2018 a modified version of the Urgent Medical Device Correction Letter and response form was sent to 23 new U.S. consignees who received Cardiosave IABP devices after the ship-hold was released. The post ship-hold devices contain the IFU addendum. This modified version of the letter notifies customers of the altitude issue and the required software update. Maquet/Getinge is currently developing a software correction to address this issue. Customers are encouraged to use the following items as specified according to altitude levels: Sensation 34cc / 40cc (Balloon Name and Size) Sensation Plus 40cc / 50cc (Balloon Name and Size) Operating Altitude: -1250 feet to 3,200 feet (795 mmHg to 676 mmHg) (1060 hPa to 901 hPa) Mega 50cc (Balloon Name & Size) Operating Altitude: -1250 feet to 5000 feet (795 mmHg to 632 mmHg) (1060 hPa to 843 hPa) Mega 30cc / 40cc (Balloon Name & Size) Linear 25cc / 34cc / 40cc (Balloon Name & Size) Operating Altitude: -1250 feet to 12,000 feet (795 mmHg to 483 mmHg) (1060 hPa to 644 hPa) For technical questions, please contact Technical Support Department (at 1-888-627-8383 and press 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST

More recalls by Maquet Datascope Corp Cardiac Assist Division

View all recalls by this company →