TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System (L.I.S)

This recall has been terminated (originally issued July 17, 2013).

Company
Technidata S.A. 387, Avenue Jean Kuntzmannn Montbonnot St Martin France
Recall Initiated
June 20, 2013
Posted
July 17, 2013
Terminated
July 29, 2013
Recall Number
Z-1743-2013
Quantity
1 site
Official Source
View on FDA website ↗

Reason for Recall

A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and

Distribution

US Distribution CA only.

Lot / Code Info

All TDHistology/Cytology product versions.

Root Cause

Software design

Action Taken

TD Technidata Medical Software sent a Medical Device Recall letter dated June 20, 2013, to one affected customer who received the TD-Synergy Histology/Cytology module. The customer was informed of the problems identified and the actions to be taken. Customers were instructed to complete the attached Reply Form and return it to the firm by fax to 514-270.2770. The customer were also instructed to contact Technidata Canada at (514) 270-7777 or at support-canada@technidata-web.com.