TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS).

This recall has been terminated (originally issued March 31, 2014).

Company
Technidata S.A. 387, Avenue Jean Kuntzmannn Montbonnot St Martin France
Recall Initiated
February 17, 2014
Posted
March 31, 2014
Terminated
March 31, 2014
Recall Number
Z-1310-2014
Quantity
85 units total (1 unit in US)
Official Source
View on FDA website ↗

Reason for Recall

If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.

Distribution

Distributed in the state of PA.

Lot / Code Info

version 11.01.A or higher

Root Cause

Software design

Action Taken

An email dated 2/17/14 was sent to their customers to inform them about concerns only if their site uses TDWorkstation and TD-Workstation Evolution products that meet all the listed criterias. The criterias are listed in the customer notification letter and it informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the reply form and return it by fax (514) 270-2770 with in 15 days from receiving the customer notification form. Customers that do not respond the reply form, Technidata will consider those customers do not wish to install the correction.