Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro diagnostic use.

This recall has been terminated (originally issued June 4, 2014).

Company
Leica Microsystems, Inc.
Recall Initiated
January 10, 2014
Posted
June 4, 2014
Terminated
September 5, 2017
Recall Number
Z-1710-2014
Quantity
145 units
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.

Distribution

Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, Romania, Russia, Saudi Arabia, Serbia, South Africa, Sweden, Taiwan, Tunisia, Turkey, and United Kingdom

Lot / Code Info

Product Code: NCL-L-CD15; Lot Numbers and Expiration Dates: 6017936, Expires 2014-08; 6015073, Expires 2014-03

Root Cause

Incorrect or no expiration date

Action Taken

Leica sent an Urgent Field Safety Correction Notice letter dated January 10, 2014 was sent ot all affected customers. The notice included instructions for recipients to: 1) destroy any unused or partially used NCL-L-CD15 (lots 6015073 and 6017936) reagent or confirm that the products are no longer in stock; 2) indicate on the attached Field Safety Correction Notice Acknowledgement Fom which action was taken; and, 3) contact the local Leica represetnative immediately if a replacement is required. Customers with any concerns about this notice can contact Leica via e-mail at LMGRA@leica-microsystems.com.