BOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use. Bond Polymer Refine Detection is a biotin-free, polymeric horseradish peroxidase (HRP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system.

Company
Leica Microsystems, Inc.
Recall Initiated
November 18, 2019
Recall Number
Z-0673-2020
Quantity
558
Firm Location
Buffalo Grove, IL

Reason for Recall

BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining.

Distribution

Nationwide and OUS to Canada.

Lot / Code Info

Lot: 65549

Root Cause

Environmental control

Action Taken

Customers were notified with an Urgent Medical Device recall letter beginning 11/18/2019. The letter identified affected product, stated reason for recall, requested that product not be used and to dispose of any product on hand. A response form was asked to be completed and returned. Questions can be directed to Adam Blunier at 1-847-821-3462 or adam.blunier@leicabiosystems.com.