RecallDepth

Hydration Station with Pedestal (Part # E30003-01), Hydration Station with Pedestal & DVD (Part # E30003-01D), Sybaritic, Inc., 9229 James Avenue South, Minneapolis, MN 55431.

Company
Sybaritic, Inc
Recall Initiated
May 15, 2007
Posted
July 15, 2009
Terminated
July 5, 2012
Recall Number
Z-1245-2009
Quantity
370
Firm Location
Bloomington, MN

Reason for Recall

Sybaritic conducted a recall on the Hydration Station 806 because of the potential for hot water to spurt through steam ports into unit and cause burns. Also, potential for customers to inadvertently place their legs over the side wall of the unit directly onto the steam ports and cause burns. There were 18 injuries and 8 malfunctions reported.

Distribution

Worldwide Distribution - US including states of AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MI, NE, NV, NH, NJ, NM, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, WV and WI and CANADA.

Lot / Code Info

Serial Numbers: 206045779, SG2061042159, 206054856, 206054874, 205024943, 206055838, 206065900, 205086311, 205104399, 205044085, 205104389, SG2061052196, SG2061242331, 205026973, 205105418, 205124542, 205085303, 206045774, SG2061042139, 206025737, SG2061252335, SG2061042144, 205036004, 206054864, 206074978, SG2070142422, 206044815, 205114444, 206045777, 205106415, 205116435, 206064932, SG2070452617, 205026946, 205105397, 205044117, SG2070352558, 206055846, 205104413, 205114480, 205104402, 206034734, 206064954, 206045775, 206055872, 206065917, 2060742003, 205024996, 206024711, 206055879, 206064931, SG2070442571, SG2061242336, SG2061052167, SG2061252330, SG2061042152, 205016864, SG2070142362, 205114443, 205105419, 206044772, 206054883, SG2061252351, 205125523, 2060942089, SG2061142232, 206064949, 206065910, 205124550, 205115440, 206045781, 205085300, SG2061152280, 2060952083, 206075982, 206045773, 206064902, SG2070152435, 205016909, 206044770, 205114464, 206044801, 206075002, SG2061252339, 206035735, 205114456, 206045778, 206055860, 2060942063, 206054889, 206044814, SG2061152278, 205126569, 206075986, SG2070142383, SG2061052140 206054842, 206034728, SG2061042175, 206054854, SG2061152250, SG2061152223, 205024974, 205014893, 206045771, 205096337, 205105400, 205025988, SG2061042164, 205026938, 205046086, 205025939, 206065964, 205025978, SG2061252345, 205104390, 206044799, SG2061252338, SG2070152353, 206055855, SG2070442574, SG2070352513, SG2070452616, SG2070452610, 206045807, 206064965, 205114494, 205105408, 205015910, 206044802, SG2061242332, SG2070342501, 206044782, SG2070452608, SG2070152378, 205116476, 205104405, 205116431, 206074983, 206064938, 205116503, 206034730, SG2061242346, 206045767, 206035723, SG2061252341, 206045776, 206065945, SG2061252329, 204115767, 206064903, 205106395, 205106398, SG2060952114, 205116506, SG2070352490, SG2061142235, 206035729, 205016896, 204126827, 206044786, 206044783, 206064913, 206044812, 205105407, SG2060952109, SG2070242441, SG2070242446, 206065953, SG2070152389, 206054888, 205014920, SG2070442573, 206044818, 206045800, SG2061252347, 206064962, 206054880, 206045822, 206064946, 205034036, 205115495, 205115495, 206055885, 206065950, SG2070152391, 206055890, 2060952076, 206045787, SG2061052213, 206044823, SG2061242327, 206065942, SG2061052160, SG2060942123, 205094380, 205114468, 206035733, 205025976, SG2070152386, 206065936, SG2070142392, SG2061242344, 20634755, SG2061042204, 206054861, 206065901, SG2070452576, SG2070152382, SG2070452572, 205126515, 205125558, SG2061042205, SG2070142356, 206074994, 206034742, 205124565, 206045813, 205104428, 206044809, 206045808, SG2070252447, 205126527, 205024977, SG2060952113, 205105422, 206055868, SG2070242472, 205115460, SG2070142385, SG2070142385, SG2060942125, 206054891, SG2061052202, 206045824, 205105392, 205024984, 206075995, 206034732, 205026981, SG2061042199, SG2060952118, 205044116, 205016898, 205106393, 206045798, 205044087, 205096350, SG2061052177, SG2060952124, SG2070152417, SG2070342556, SG2070452622, SG2061252343, 204095524, SG2070142373, SG2070152407, SG2070142354, 206045817, 206065951, 206065963, SG2061242334, 205024937, 206045819, 2060655873, SG2070152364, SG2061042169, SG2060942119, SG2060952120, SG2061142269, SG2060942108, SG2060942105, 206054865, SG2070242440, 205114484, 205016923, SG2070142396, SG2070152399, 205084307, 206034721, SG2061042151, SG2061042171, SG2061152283, SG2061052168, 204095422, 206054850, 206065909, 206035724, 206045810, 205035068, SG2070142433, 206045784, 204125834, SG2070542669, 206055884, 206044820, SG2070152393, SG2061052161, 206055886, SG2061052172, SG2070152432, 206065944, 206034750, 206055859, SG2070142371, 205015895, 205024972, SG2070252454, SG2070252469, SG2070342551, SG2070352492, SG2070352552, 204096525, 204116734, SG2061252349, SG2070142355, 205106862, SG2060952115, 206074979, 206035727, 205114436, SG2061052141, 206064943, SG2070142411, SG2061252340, SG2070142434, SG2070352553, 206055858, SG2061042197, SG2061242350, 205046089, 205115493, 205014863, 206044785, SG2070152399, SG2070142394, SG2070252467, SG2070242481, SG2070342557, 205036066, SG2070152366, 205014894, SG2061052166, 2060942082, SG2070152395, SG2070152358, 206045805, SG2070442634, SG2070252444, SG2061042155, SG2061052176, 205116432, 205116511, 206026670, 206054887, 206075998, SG2060942104, SG2070252471, 206036720, SG2060942084, 205045088, 205124576, SG2061242333, 204116738, SG2061052150, SG2061052203, 205036042, SG2070152378, SG2070242482 and SG2070242488.

Root Cause

Device Design

Action Taken

Sybaritic, Inc. issued a letter dated May 15, 2007 addressed to "Hydration Station Customer". The letter informed the customer of the new 2007 Hydration Station Upgrade Improvement Kits including instructions and advised the customer to coordinate the installation of the components with their Technical Service Department. Further questions can be addressed to Sybaritic, Inc. Technical Administration at 1-800-445-8418 extension 373.