DERMALIFE Spa Jet 2G System. Part #E30006-01 - Spa Jet 2G; Assay w/Pedestal. This is the Spa Jet product including the 2 optional pedestals which the unit can sit on top of. Part #E30006 - Spa Jet 2G; 220V,50/60Hz. 15A. This is the Spa Jet product without the 2 optional pedestals which the unit can sit on top of.
- Company
- Sybaritic, Inc
- Recall Initiated
- June 22, 2007
- Posted
- July 13, 2009
- Terminated
- July 5, 2012
- Recall Number
- Z-1602-2009
- Quantity
- 238 units
- Firm Location
- Bloomington, MN
Reason for Recall
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Sybaritic, Inc. conducted a recall on DERMALIFE 2G equipment (SpaJet, Spa Oceana, Spa Feng Shui) because of IR Fan failure which resulted in the overheating and failure of the IR emitter and surrounding assembly.
Distribution
Worldwide Distribution -- US including states of CA, IA, MN, NE, NJ, NY, TX and WA and countries of Macedonia, Hungary, Greece, South Africa, Italy, Hong Kong, United Arab Emirates, Greece, Croatia, Australia, Kuwait, Turkey, Mexico, Belgium, England, Barbados, France, Japan, Lebanon, Moldova, Poland, Portugal, Romania, Russia, Spain and Thailand.
Lot / Code Info
PART E30006 SERIAL NUMBERS: G1070442568, G1070452567, G1060942061, G1061252308, G106064921, G106065928, G10606891, G1061142252, G1061152287, G1061152288, G1061242302, G1070352519, G1061142238, G1061142247, G1061142270, G1061152265, G1061252301, G1061252306, G1070342543, G1060942112, G1061142292, G1061152267, G1070142402, G1070242450, G1070252451, G1070342515, G1070342524, G1070342527, G1070352537, G1070342554, G1070352562, G1070442582, G1070452591, G106075989, G1061042162, G1061042189, G1061052180, G1061052183, G1061052185, G1061052193, G1070542653, G106125312, G1060852039, G1061142239, G1061152237, G1061152251, G1070342508, G1070342511, G1070352514, G1070352555, G1070452580, G1070452586, G1040452600, G106034717, G106054836, G106055876, G106055877, G106055881, G106065924, G106074984, G106074990, G106075988, G106075992, G1060842008, G1060842010, G1060842015, G1060842041, G1060852012, G1060852051, G1060942072, G1060942111, G1060942116, G1060952064, G1060952065, G1060952087, G1060952117, G1060952122, G1061042134, G1061042179, G1061042184, G1061042187, G1061042191, G1061042208, G1061052132, G1061052133, G1061052138, G1061052142, G1061052143, G1061052147, G1061052154, G1061052163, G1061052188, G1061052194, G1061052207, G1061052215, G1061142218, G1061142246, G1061142262, G1061142264, G1061142276, G1061152222, G1061152233, G1061152241, G1061152242, G1061152253, G1061152254, G1061152256, G1061152272, G1061152274, G1061152297, G1061152298, G1061242300, G1061242305, G1061242314, G1061242315, G1061252319, G1061252320, G1070142374, G1070142375, G1070142377, G1070142397, G1070142403, G1070142410, G1070142413, G1070142415, G1070142419, G1070142427, G1070152360, G1070152361, G1070152368, G1070152388, G1070152418, G1070152420, G1070152424, G1070152425, G1070152426, G1070152428, G1070242452, G1070242470, G1070242480, G1070252443, G1070252449, G1070252453, G1070252478, G1070252479, G1070252485, G1070342494, G1070342499, G1070342510, G1070342531, G1070342535, G1070342539, G1070342541, G1070342559, G1070342561, G1070352489, G1070352496, G1070352497, G1070352500, G1070352504, G1070352530, G1070352538, G1070352546, G1070352560, G1070442565, G1070442569, G1070442597, G1070452563, G1070452564, G1070452566, G1070452579, G1070452583, G1070452594, G1070542654, G1060952107, G1070142398, G1070152401, G1070552671, G1070552684, G106074991, G1061042181, G1061042198, G1061052209, G1061052210, G1061052216, G1061142217, G1061142226, G1061142231, G1061142243, G1061142245, G1061152221, G1061152229, G1061152230, G1061152234, G1070142423, G1070242475, G1070342503, G1070342525, G1070352507, G1070352520, G1070352522, G1070352523, G1070352526, G1070352534, G106044763, G106045769, G106054827, G106054844, G106054848, G106054878, G106055834, G106055840, G106055841, G106055881, G106054875, G106054896, G106055893, G106055894, G106054882, G106064923, G106064927, G106065933, G406055869 and G106054870; and PART E30006-01 SERIAL NUMBERS: G1060852047, G1061142255, G1061152260, G1061152261, G1061152271, G1061152275, G1060852059, G1060842045, G1060852034, G1061242321, G1061252313, G106852043 and G1060852009.
Root Cause
Device Design
Action Taken
Sybaritic, Inc. issued a "Safety Alert" dated April 15, 2008 informing "All Owners/Users of DermaLife 2G Equipment" of the affected device. The letter advised consignees on device use until the mandatory Filed Correction Kit is delivered. For further questions, contact Sybaritic, Inc. at 1-800-445-8418.