Boston Scientific LeVeen SuperSlim Needle Electrode, 2.0 cm/25 cm, sterile. The device is intended to be used in conjunction with a Radio Therapeutics Corporation radiofrequency (RF) generator for thermal coagulation necrosis of soft tissue, including partial or complete ablation of nonresectable liver lesions.

This recall has been terminated (originally issued September 16, 2008).

Company
Boston Scientific Corp
Recall Initiated
October 1, 2007
Posted
September 16, 2008
Terminated
September 16, 2008
Recall Number
Z-2092-2008
Quantity
15,604 of all products
Firm Location
Spencer, IN
Official Source
View on FDA website ↗

Reason for Recall

The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.

Distribution

Japan.

Lot / Code Info

All lots.

Root Cause

Device Design

Action Taken

Customer notifications were sent to customers in Japan on October 1, 2007 instructing them to check their inventory and to return affected product. Contact Rob Miragliuolo at 508-683-4186 if you have questions.