Boston Scientific 1.0mm x 570 mm rigid pneumatic probe for use only with Swiss LithoClast Select System model no. 840-300K, Swiss LithoClast Ultra System model nol 840-700K, or Swiss LithoClast Ultra Upgraded System model no. 840-701U, compatible with 330mm Ultrasound Probes (REF 840-714 and 840-715), Rx, sterile, no detectable latex, Packaged and Distributed by Boston Scientific Corporation, Natick, MA; REF/catalog no 840-733.

This recall has been terminated (originally issued November 11, 2009).

Company
Boston Scientific Corp
Recall Initiated
September 17, 2009
Posted
November 11, 2009
Terminated
October 8, 2010
Recall Number
Z-0175-2010
Quantity
173
Firm Location
Spencer, IN
Official Source
View on FDA website ↗

Reason for Recall

The label incorrectly reads "compatible with 330mm ultrasound probes (Ref 840-714 and 840-715)" instead of bearing the correct statement "compatible with 403mm ultrasound probes (Ref 840-716 and 840-7170."

Distribution

Nationwide.

Lot / Code Info

Lots 12723202, 12745667, 12767883, 12777930, 12782625, 12808559, 12813607, 12850772 and 12861508.

Root Cause

Other

Action Taken

Consignees were notified by letter titled "Urgent Medical Device Recall-Immediate Action Required" dated 9/17/09 and instructed to cease using the product and to call the firm's customer service for return authorization. For additional information, call the firm at 866-868-4004.