Hospira Monitoring Kit (PA Line) with 03ml Flush Device for Saginaw General Hosp. , Item Number: 460760451 Indicated for fluid flow directional control and for providing access port(s) for administration of solutions.

This recall has been terminated (originally issued September 16, 2008).

Company
Icu Medical, Inc.
Recall Initiated
June 11, 2008
Posted
September 16, 2008
Terminated
April 13, 2010
Recall Number
Z-2206-2008
Quantity
120
Firm Location
Amanecer San Clemente, CA
Official Source
View on FDA website ↗

Reason for Recall

Improper orientation of the stopcock handle questions whether there is an adequate gas path to assure sterility, which presents a possible compromise of sterility.

Distribution

Nationwide Distribution

Lot / Code Info

Lot Number: 41-501-LA

Root Cause

Process change control

Action Taken

Every customer both direct and distributors who have purchased the affected lots were mailed a Product Recall Letter with attached fax-back Product Inventory Summary attachment on 06/11/2008 via fax and followed by UPS overnight delivery. The Product Recall Letter informed the customers of the reason for the recall and instructed them to complete and return the attached Product Inventory Summary. Instructions were also given for returning product in their inventory to ICU.