GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube, collimator and image intensifier; wall stand; Overhead tube suspension; operator console and digital archive system.

This recall has been terminated (originally issued September 22, 2006).

Company
General Electric Med Systems Llc
Recall Initiated
May 15, 2006
Posted
September 22, 2006
Terminated
May 25, 2007
Recall Number
Z-1304-06
Quantity
1,289
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

Systems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in position to intercept the entire cross section of the useful beam

Distribution

Worldwide, including USA, Puerto Rico, Canada, France, Germany, Indonesia, Italy, Japan, Jordan, Kuwait, Macedonia, Panama, Russia, Saudi Arabia, Singapore, and United Kingdom.

Lot / Code Info

All serial numbers affected

Root Cause

Other

Action Taken

An Important Notice for Precision 500D, dated 11/24/2005 was hand delivered to all customers beginning 05/15/06, by GE field service engineers at the time they were to perform the correction to the device.