RecallDepth

GE Proteus XR/a Radiographic System. GE Medical Systems-Americas Milwaukee, WI. Model Number: 2259988. Product is indicated for use in generating radiographic images of human anatomy in all general-purpose diagnostic procedures. This device is not intended for mammographic applications.

This recall has been terminated (originally issued June 8, 2012).

Company
General Electric Med Systems Llc
Recall Initiated
May 31, 2006
Posted
June 8, 2012
Terminated
July 31, 2012
Recall Number
Z-1790-2012
Quantity
3,796 units
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath

Distribution

Worldwide Distribution.

Lot / Code Info

All serial numbers

Root Cause

Device Design

Action Taken

GE Healthcare sent an "URGENT SAFETY INFORMATION" letter dated May 31, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included new operator instructions and warning labels.

More recalls by General Electric Med Systems Llc

View all recalls by this company →