AcuSnare Polypectomy Device, Cook Wilson-Cook Medical GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105. The SASM device prefix denotes Soft AcuSnare Mini. The ASM device prefix denotes Acu Snare Mini. Due to the nitinol core wire, the difference between these two products is stiffness of the snare head wire.

Company
Wilson-Cook Medical Inc
Recall Initiated
April 21, 2004
Posted
July 20, 2004
Terminated
June 10, 2004
Recall Number
Z-0906-04
Quantity
5 boxes/10 pouches per box
Firm Location
Winston Salem, NC

Reason for Recall

The snares inside the box labeled SASM-1 may actually be ASM-1 devices. The individual pouches may be correctly labeled ASM.

Distribution

MA, NY, OK

Lot / Code Info

Reorder (Catalog) number: ASM-1, Lot#W1834120 and Reorder (Catalog) number: SASM-1, Lot#W1834310

Root Cause

Other

Action Taken

Consignees were notified by visit by sales representatives with follow up letter on 4/21/2004.