Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. It is indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.

Company
Wilson-Cook Medical Inc
Recall Initiated
December 18, 2025
Posted
February 13, 2026
Recall Number
Z-1362-2026
Quantity
153
Firm Location
Winston Salem, NC

Reason for Recall

Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.

Distribution

U.S. Distribution to states of: CA, CO, IN, and NC.

Lot / Code Info

RPN: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 Work Order Number / RPN W4954835 / TESLA-B2535-260 W4954836 / TESLA-B2535-260 W4957208 / TESLA-B2535-260 W4958992 / TESLA-B2535-260 W4960895 / TESLA-B2535-260 W4960896 / TESLA-B2535-260 W4961670 / TESLA-B2535-260 W4965708 / TESLA-B2535-260 W4966040 / TESLA-B2535-260 W4966803 / TESLA-B2535-260 W4967866 / TESLA-B2535-260 W4968515 / TESLA-B2535-260 W4968867 / TESLA-B2535-260 W4969292 / TESLA-B2535-260 W4969674 / TESLA-B2535-260 W4969687 / TESLA-B2535-260 W4970315 / TESLA-B2535-260 W4970662 / TESLA-B2535-260 W4971561 / TESLA-B2535-260 W4972693 / TESLA-B2535-260 W4973085 / TESLA-B2535-260 W4974435 / TESLA-B2535-260 W4974588 / TESLA-B2535-260 W4975407 / TESLA-B2535-260 W4975688 / TESLA-B2535-260 W4977514 / TESLA-B2535-260 W4977824 / TESLA-B2535-260 W4979182 / TESLA-B2535-260 W4979183 / TESLA-B2535-260 W4979665 / TESLA-B2535-260 W4982425 / TESLA-B2535-260 W4982436 / TESLA-B2535-260 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 Work Order Number / RPN: W4962311 / TESLA-B2535-450 W4966039 / TESLA-B2535-450 W4968222 / TESLA-B2535-450 W4969306 / TESLA-B2535-450 W4970322 / TESLA-B2535-450 W4971499 / TESLA-B2535-450 W4975687 / TESLA-B2535-450 W4977254 / TESLA-B2535-450 W4982013 / TESLA-B2535-450

Root Cause

Under Investigation by firm

Action Taken

On 12/18/2025, the firm priority mailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that the firm is removing Teslatome Bipolar Sphincterotomes from the market because they are nonconforming. Customers are instructed to: 1. Examine inventory immediately to determine if you have affected devices. Immediately cease use of the affected devices and quarantine all affected devices. 2. Return the affected devices to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a credit. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected devices on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. For questions, contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235