Leica M822 Surgical Microscope Device Controller Software (MDC version 3.6, pack F, affecting the control unit component, Article #10448446, in combination with a Leica M822 Optics Carrier-XY-Focus-Tilt-Unit, article #10448587; Leica Microsystems (Schweiz) AG, Max Schmidheiny-Strasse 201, CH-9425, Heerbrugg, Switzerland The intended use of the microscope is for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

This recall has been terminated (originally issued February 16, 2012).

Company
Leica Microsystems, Inc.
Recall Initiated
December 21, 2011
Posted
February 16, 2012
Terminated
February 16, 2012
Recall Number
Z-1042-2012
Quantity
29 units (10 units-USA)
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

Leica Microsystems received complaints stating that the zoom function of the Leica M822 surgical microscope locked up during use.

Distribution

Worldwide distribution: USA (nationwide) including states of: California, Colorado, Illinois, New York, Oregon, South Carolina; and countries of: Australia, Japan, Korea, Pakistan, Romania, South Africa, Switzerland and Turkey.

Lot / Code Info

Article 10448446, serial numbers 181011001, 181011009, 181011005, 181011003, 250811007, 28101108, 151111011, 211111007, 251111006 and 251111004. The corrected units will have controller software version MDC V3.6 pack G

Root Cause

Device Design

Action Taken

The firm, Leica Microsystems, Inc., sent a "Medical Device Correction" notice dated December 21, 2011 to all affected customers by e-mail. The product, problem, and actions to be taken by the customers were addressed. Customers were instructed- DO NOT USE use the affected product until the software could be upgraded and complete and return the Acknowledgement Form via fax to: 1-847-607-3147 or Scan and send to elizabeth.culotta@leica-microsystems.com. Leica Microsystems representatives will contact the customers to make necessary arrangements to complete the software upgrade. If you have any questions, contact Director, Global Product Quality Management at 1-847-317-7209.