Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Within each product family, there are varying features, rate responsive pacing, and diagnostic data. Medtronic,Kappa 700 KD700,KD701,KD703,KD706,KDR700,KDR700V,KDR701,KDR701PTO,KDR701V,KDR701VIT,KDR703,KDR703V,KDR706,KDR706V,KDR720,KDR721,KDR730,KDR731,KDR733,KVDD700,KV DD701 Medtronic,Kappa 800 KDR801,KDR803,KDR806 Medtronic,Kappa 900 KD901,KD903,KD906,KDR901,KDR903,KDR906,KDR921,KDR931,KDR933,KVDD901 Medtronic, Enpulse DR E1DR01,E1DR06,E1DR21,E2D01,E2D03,E2DR01,E2DR03,E2DR06,E2DR06J,E2DR21,E2DR21PJ,E2DR31,E2DR31J,E2DR33,E2DR33J,E2VDD01 Medtronic, Adapta/Versa/Sensia DR ADD01,ADDR01,ADDR03,ADDR06,ADDRL1,ADDRS1,ADVDD01,SED01,SED01N,SEDR01,SEDR01N,SEDRL1,VEDR01 Medtronic, Relia RED01,REDR01,REVDD01 (approved in the US but has not been sold in US) Medtronic, Vitatron Extension G and E series DR E50A1,E60A1,G70A1 (not sold in the US)

This recall has been terminated (originally issued April 29, 2011).

Company
Medtronic Inc. Cardiac Rhythm Disease Management
Recall Initiated
April 4, 2011
Posted
April 29, 2011
Terminated
November 27, 2012
Recall Number
Z-2120-2011
Quantity
1,989,180
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

A rare measurement lock-up condition that may (at a rate of approximately 1 in 18,000 devices) inappropriately trigger ERI/RRT in the Medtronic Dual Chamber Pacemakers: Kappa 600, 700, 800, 900, EnPulse, Adapta, Versa, Sensia. This is caused by a random lock-up of the measurement system hardware that may result in an incorrect battery voltage reading of zero. This issue does NOT impact battery lon

Distribution

Worldwide distribution: USA (nationwide) including Guam, Puerto Rico and DC; and countries including: Algeria, Andorra, Anguilla, Argentina, Armenia, Aruba, Australia, Austria, Bahamas, Bangladesh, Belgium, Bermuda, Bolivia, Brazil, Brunei, Darussalam, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Virgin Islands (British).

Lot / Code Info

All serial numbers affected.

Root Cause

Component design/selection

Action Taken

The firm, Medtronic, sent an "Important: Medical Device Correction" letter dated April 2011 to its customers. The letter described the product, problem, and actions taken. Medtronic informed the customers that this issue does NOT impact battery longevity and does NOT require device explant. Currently, the device can be reset to normal operation by a Medtronic representative. Reset devices are no more likely to experience a recurrence of this issue. Medtronic is planning to release a programmer software update that will allow the clinician to reset the device. In addition, Medtronic has consolidated their product performance data with the vision of creating a primary reference tool for performance updates. If you have any questions, please contact your local Medtronic Representative or Medtronic Technical Services at 800-505-4636.