Artisan brand Extend Control Catheter; An Steerable Guide Catheter; Component of Sensei X Robotic Catheter System; Model Number: 07546; Distributed and Manufactured by Hansen Medical, Inc, Mountain View, CA

This recall has been terminated (originally issued April 6, 2010).

Company
Hansen Medical Inc
Recall Initiated
October 19, 2009
Posted
April 6, 2010
Terminated
February 9, 2011
Recall Number
Z-1251-2010
Firm Location
Mountain View, CA
Official Source
View on FDA website ↗

Reason for Recall

Certain Artisan Extend Catheters may develop a leak in the flush assembly of the catheter which has the potential to cause flush fluid leakage and subsequent decrease in distal flush; inability to aspirate air from flush lines, and/or introduction of air into the catheter.

Distribution

Nationwide to 12 states.

Lot / Code Info

Serial Numbers:  6394, 6376, 6377, 6378, 6379, 6380, 6381, 6382, 6383, 6384, 6385, 6386, 6387,  6388, 6389, 6390, 6392, 6393, 6395, 6396, 6397, 6398, 6399, 6400, 6401, 6402,  6403, 6406, 6407, 6408, 6409, 6410, 6411, 6412, 6413, 6414, 6415, 6417, 6418,  6419, 6420, 6421, 6422, 6423, 6424, 6425, 6427, 6428, 6429, 6430, 6431, 6432,  6433, 6434, 6435, 6436, 6437, 6438, 6439, 6440, 6441, 6442, 6443, 6444, 6445,  6446, 6447, 6448, 6449, 6450, 6451, 6452, 6453, 6454, 6455, 6456, 6457, 6458,  6459, 6460, 6461, 6462, 6463, 6464, 6465, 6466, 6467, 6468, 6469, 6470, 6471,  6472, 6473, 6474, 6475, 6476, 6477, 6478, 6479, 6480, 6481, 6482, 6483, 6484,  6485, 6486, 6487, 6488, 6489, 6490, 6491, 6492, 6493, 6494, 6495, 6496, 6497,  6498, 6499, 6501, 6502, 6504, 6505, 6506, 6507, 6508, 6509, 6511, 6513, 6514,  6515, 6516, 6517, 6519, 6520, 6522, 6523, 6524, 6525, 6526, 6527, 6528, 6530,  6532, 6533, 6534, 6537, 6542, 6543, 6544, 6549, 6556, 6562, 6564, 6566, 6569,  6570, 6573, 6574, 6575, 6577, 6578, 6580, 6582, 6585, 6586, 6588, 6589, 6591,  6593, 6596, 6598, 6599, 6600, 6601, 6608, 6612, 6615, 6616, 6620, 6621, 6626,  6627, 6633, 6635, 6636, 6638

Root Cause

Other

Action Taken

Recall Initiation Date: October 19,2009 (FedEx shipments on October 19, for October 20 delivery) via letters explaining the reason for the recall and requesting the product be returned. Customers with questions are instructed to call the company at 650-404-5940.