Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

This recall has been terminated (originally issued April 15, 2015).

Company
Hansen Medical Inc
Recall Initiated
March 18, 2015
Posted
April 15, 2015
Terminated
September 23, 2015
Recall Number
Z-1449-2015
Quantity
19 total installed systems
Firm Location
Mountain View, CA
Official Source
View on FDA website ↗

Reason for Recall

Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.

Distribution

Worldwide Distribution - US Distribution and the countries of : Saudi Arabia, Australia, France, Italy, Germany, and United Kingdom.

Lot / Code Info

All serial numbers.

Root Cause

Software design

Action Taken

Hansen Medical sent an Urgent Notice - Field Safety Notice dated March 24, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. We request that you review the information in pages 1 and 2 with Hansen Medical personnel and acknowledge your understanding by signing the Field Safety Notice Training Form on page 3. If you need any further information regarding this issue, please do not hesitate to contact me via telephone at 650-404-2784.