RecallDepth

METRx II Tube, 22mm x 6cm stainless steel tubular retractor, Part No. 9569733

This recall has been terminated (originally issued April 12, 2006).

Company
Medtronic Sofamor Danek Instrument Manufacturing
Recall Initiated
January 25, 2006
Posted
April 12, 2006
Terminated
July 5, 2007
Recall Number
Z-0750-06
Quantity
2
Firm Location
Bartlett, TN
Official Source
View on FDA website ↗

Reason for Recall

Due to a manufacturing error after autoclaving the Tubular Retractor coating, Physical Vapor Deposition (PVD) black Titanium Nitride (TiN), may rub off.

Distribution

IL, IN, MI, CA, Japan, Netherlands

Lot / Code Info

Lot No. DC05L008

Root Cause

Other

Action Taken

The firm sent out a recall letter dated January 25, 2006 stating the problem and requesting the return of all unused product.