METRx II Tube, 14mm x 5cm stainless steel tubular retractor, Part No. 9569662
This recall has been terminated (originally issued April 12, 2006).
- Recall Initiated
- January 25, 2006
- Posted
- April 12, 2006
- Terminated
- July 5, 2007
- Recall Number
- Z-0721-06
- Quantity
- 1
- Firm Location
- Bartlett, TN
- Official Source
- View on FDA website ↗
Reason for Recall
Due to a manufacturing error after autoclaving, the Tubular Retractor coating, Physical Vapor Deposition (PVD) black Titanium Nitride (TiN), may rub off.
Distribution
IL, IN, MI, CA, Japan, Netherlands
Lot / Code Info
Lot No. FA05K005
Root Cause
Other
Action Taken
The firm sent out a recall letter dated January 25, 2006 stating the problem and requesting the return of all unused product.