LOCIQ Book System models: 2399921, 2399923, 2349933, 2349934 and 2349935
This recall has been terminated (originally issued January 11, 2006).
- Company
- General Electric Med Systems Llc
- Recall Initiated
- September 2, 2004
- Posted
- January 11, 2006
- Terminated
- January 5, 2007
- Recall Number
- Z-0387-06
- Quantity
- 2698
- Firm Location
- Waukesha, WI
- Official Source
- View on FDA website ↗
Reason for Recall
The Estimated Fetal Weight (EFW) result will be incorrect if the operator fails to perform the 'end exam' or new patient registration step before starting an EFW measurement procedure on the next patient.
Distribution
nationwide and worldwide
Lot / Code Info
all serial numbers
Root Cause
Other
Action Taken
Devices will have software upgrades done by GE Field Service Technicians. The upgrade will include a dialog box which will display on the screen when prior data exists and a new patient is being registered.