RecallDepth

IMMULITE 2000 Vitamin B12 Kit, Catalog # L2KVB2, L2KVB6, kit lot 186.

Company
Diagnostic Products Corp.
Recall Initiated
January 24, 2005
Posted
February 11, 2005
Terminated
July 12, 2005
Recall Number
Z-0537-05
Quantity
189 kits
Firm Location
Los Angeles, CA

Reason for Recall

A low CPS limit was not incorporated in the kit barcode for IMMULITE 2000 Vitamin B12 L2KVB 2, 6 kit lot 186 only.

Distribution

AL, CA, CO, FL, IL, KS, LA, NY, MA, MD, MI, MS, NC, NE, NM, OH, PA, TN, TX, UT, WA, and Australia, Czech Republic, France, Germany, Netherlands, Portugal, Puerto Rico, Sweden, Turkey, UK, Venzuela

Lot / Code Info

IMMULITE 2000 Vitamin B12 Kit, Catalog # L2KVB2, L2KVB6, kit lot 186.

Root Cause

Other

Action Taken

Domestic customers were notified on January 24, 2005, via telephone and strted receiving a fax-back form and a technical Bulletin on January 25, 2005. Foreign consignees were notified via email and/or Fax.

More recalls by Diagnostic Products Corp.

View all recalls by this company →