IMMULITE 2000 Total Testosterone Kit

Company
Diagnostic Products Corp.
Recall Initiated
January 3, 2006
Posted
February 16, 2006
Terminated
August 30, 2006
Recall Number
Z-0528-06
Quantity
710
Firm Location
Los Angeles, CA

Reason for Recall

The reagent for this kit lot (L2KTW 2,6 126) was manufactured with a raw material that can carryover and interfere with DPC Kit IMMULITE 2000 Estradiol (L2KE2) causing a false elevation in the observed estradiol concentration.

Distribution

Nationwide and Worldwide in Pakistan, Italy, Cyprus, Turkey, Switzerland, Chile, South Africa, Greece, Mexica, Canada, Thialand, Germany, India, United Arab Emirates, Hungary, Israel, Portugal, Taiwan, Austria, Saudi Arabia, Malaysia, Hong Kong, India, Colombia, Argentina, Australia, Costa Rica, Czech Republic, Spain, France, Panama, Brazil, Netherlands, Poland, Sweden, United Kingdom, & El Salvadon.

Lot / Code Info

CATALOG & LOT NUMBERS: L2KTW 2 Lot 126 (200-test kit) L2KTW 6 Lot 126 (600-test kit)

Root Cause

Other

Action Taken

For domestic customers - the recall is extended to the domestic end user-level. Domestic customers were notified via telephone using a Phone Script (attached in Attachment 5), starting on January 5,2005. Each customer receives a fax-back form and a Technical Bulletin directly after this phone contact. For foreign consignees - the recall is extended to foreign distributors. It is the distributor''s responsibility to extend the recall action to their direct customers. Foreign consignees were notified via e-mail and/or FAX,starting on January 3, 2005.