Oridion Medical Capnostream20 capnograph/pulse oximeter monitor and monitor with printer. Part numbers CS 08651, CS 08653, CS08657, and CS08659.

This recall has been terminated (originally issued September 21, 2009).

Company
Oridion Medical
Recall Initiated
June 30, 2009
Posted
September 21, 2009
Terminated
October 1, 2010
Recall Number
Z-2232-2009
Quantity
326 units
Official Source
View on FDA website ↗

Reason for Recall

1) The monitor's ability to withstand electrostatic discharge (ESD) to exposed metal connectors on the back of the monitor which may cause the monitor's display to freeze. 2) The unit generates electromagnetic radiation above the limits set by applicable standards for medical devices

Distribution

Nationwide.

Lot / Code Info

Software version V 3.0 and earlier.   Part Number Part Description   CS08651, Capnostream20 with Nellcor pulse oximetry module   CS08653 Capnostream20 with Masimo pulse oximetry module   CS08657 Capnostream20 with Nellcor pulse oximetry module and printer   CS08659  Capnostream20 with Masimo pulse oximetry module and printer

Root Cause

Device Design

Action Taken

Oridian sent a letter to all its distributors June 25, 2009, instructing them to gather all the units to be exchanged for factory reconditioned units that have the hardware and software enhancements and meet all performance requirements in full.