Oridion Surestream CO2 Sampling Lines- Smart SureLine Plus O2 adult Part Number: 010981-03 Product Usage: The intended use of SureStream sampling lines is to convey a sample of a patients expired breath to a capnograph in order to measure CO2 partial pressure in intubated and non intubated patients.

This recall has been terminated (originally issued March 30, 2012).

Company
Oridion Medical
Recall Initiated
December 1, 2011
Posted
March 30, 2012
Terminated
June 4, 2012
Recall Number
Z-1347-2012
Quantity
70 boxes (1750 unts)
Official Source
View on FDA website ↗

Reason for Recall

Surestream CO2 Sampling Lines-Connector may cause breakage of the microswitch inside the CO2 inlet

Distribution

USA Nationwide Distribution including the states of: AZ, CA, FL, HI, LA, MA, MO, NJ, NY, OH, OR, TX and WI.

Lot / Code Info

Lot Numbers: M8209L10, M8262M10

Root Cause

Device Design

Action Taken

Oridion issued a Technical Note communication in March 2011 to field staff and end users globally. The Technical Note identified the affected product and provided instructions for user on the proper insertion of Surestream products. For questions contact your local distributor or Oridion at technicalsupport@oridion.com