Biosense Webster PREFACE Guiding Sheath, Multipurpose 77cm, for the introduction of intravascular electrophysiology catheters into any cardiac chamber. Catalog Number 301805M

Company
Biosense Webster, Inc.
Recall Initiated
November 17, 2006
Posted
February 28, 2007
Terminated
June 25, 2009
Recall Number
Z-0539-2007
Quantity
996
Firm Location
Irwindale, CA

Reason for Recall

Biosense Webster, Inc. has received a number of complaints regarding the radiopaque tip separating or partially detaching from the PREFACE Sheath during use in the left atrium.

Distribution

Worldwide distribution, including USA, Australia, Brazil, Canada, China, Colombia, Hong Kong, Indonesia, Israel, India, Japan, Korea, the Netherlands, New Zealand, Pakistan, Singapore, Thailand, and Taiwan.

Lot / Code Info

Lot numbers: 13073872,  13078568,  13082798,  13084602,  13088356,  13096358,  40206984, 40604122, 40804442,  50205751, 50306355,  50403002,  50405034,  51104857,  A0105904,  A0106063,  A0206419, A1102924, X0105497,  X0205439,  X0305391,  X0306127, X0306956,  X0405043,  X0505118,  X0704423, X0802743, X0805081 & X0905452

Root Cause

Other

Action Taken

Initially a Medical Device Correction Advisory letter was mailed on November 20, 2006 to all domestic and international consignees recommending to customers that the Preface Sheath was not to be used with the NaviStar ThermoCool Catheter or any 8 French or larger catheter. On December 9, 2006, a Recall Customer letter and Certification form were sent out to all customers and Johnson & Johnson affiliates who had purchased the product in the past three years. For customers in the US, two letters and forms were sent out -one addressed to the EP/Cardiology Lab and the second to the Risk Management Office. For the US, customers will return product directly to Biosense Webster afler receiving a return authorization number. Returned product will be segregated and labeled as quarantined product. Returned product will be used as necessary for the root cause investigation. Any product not used at the conclusion of the investigation will be destroyed. For international customers, product will be returned to the Biosense Webster warehouse in Roden, Netherlands or to the Biosense Webster US warehouse (depending on order placement) if export issues don''t present a problem. If product cannot be easily returned to Biosense Webster, the customers will be instructed to destroy the product and provide notification that the destruction was performed.