NicoletOne version 5.20 software (also known as NicoletOne Version 5.2 Software) used with the NicoletOne LTM, Sleep, nEEG, vEEG and ICU monitoring Systems.

This recall has been terminated (originally issued February 15, 2007).

Company
Nicolet Biomedical Div Of Viasys Healthcare
Recall Initiated
August 25, 2006
Posted
February 15, 2007
Terminated
December 10, 2011
Recall Number
Z-0380-2007
Quantity
75
Firm Location
Madison, WI
Official Source
View on FDA website ↗

Reason for Recall

A software anomaly exists in the 5.20.1038 NicoletOne LTM, Sleep, nEEG, vEEG and ICU Monitor systems using the M, and/or C series amplifiers. If this anomaly occurs, the system will display the data for channel one in all channels on NicoletOne LTM, Sleep, nEEG, vEEG or ICU Monitor systems.

Distribution

Worldwide, including USA, Belgium, Chile, Denmark, Italy, Japan, Poland, and Spain.

Lot / Code Info

NicoletOne Version 5.20 software released and installed after August 07, 2006.

Root Cause

Other

Action Taken

An Advisory Notice, dated August 25, 2006, was sent to all affected customers. The notice described the issues, states to discontinue use of 5.20.1038 software until updated 5.21.1039c software is installed.