RecallDepth

Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711

This recall has been terminated (originally issued January 3, 2008).

Company
Nicolet Biomedical Div Of Viasys Healthcare
Recall Initiated
September 11, 2007
Posted
January 3, 2008
Terminated
December 10, 2011
Recall Number
Z-0126-2008
Quantity
33
Firm Location
Madison, WI
Official Source
View on FDA website ↗

Reason for Recall

Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.

Distribution

Worldwide Dsitribution - USA, Germany, Netherlands, Kuwait, Japan, Ireland, and China.

Lot / Code Info

ON040060, ON050120, ON050130, ON050168, ON050186, ON050197, ON050199, ON050203, ON050206, ON050212, ON050220, ON050231, ON060232, ON060243, ON060254, ON060256, ON060257, ON060269, ON060270, ON060272, ON060275, ON060276, ON060277, ON060278, ON060279, ON060280, ON060288, ON060290, ON060291, ON060292, ON060299, ON070301, ON070308.

Root Cause

Component design/selection

Action Taken

Certified direct mail delivery of URGENT MEDICAL DEVICE FIELD CORRECTION (RECALL) were made to consignees on 9/11/07 by Nicolet Biomedical Div of Viasys Healthcare. The letter describes the product and problem. VIASYS requests that customers cease use of the video option of the device until the monitor can be replaced by VIASYS authorized personnel. Letter also requests consignees to fax or mail back the response form.

More recalls by Nicolet Biomedical Div Of Viasys Healthcare

View all recalls by this company →