MICRO-X Rover Mobile X-ray System, # MXU-RV35

This recall is currently active, issued December 12, 2024. It was issued by Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia.

Company
Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia
Recall Initiated
May 31, 2023
Posted
December 12, 2024
Recall Number
Z-0664-2025
Quantity
5
Official Source
View on FDA website ↗

Reason for Recall

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Distribution

Worldwide and US Nationwide Distribution: NJ, PR, CA

Lot / Code Info

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Root Cause

Software design

Action Taken

The firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.