RecallDepth

Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A

This recall is currently active, issued April 24, 2025. It was issued by Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia.

Company
Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia
Recall Initiated
April 4, 2025
Posted
April 24, 2025
Recall Number
Z-1670-2025
Quantity
5 units
Official Source
View on FDA website ↗

Reason for Recall

Potential for internal fastening mechanism within generator of mobile x-Ray to fail.

Distribution

US: NJ, WA, Puerto Rico OUS: None

Lot / Code Info

Serial Number/UDI-DI - 349 (01)09357123000037(11)230209(21)00349 348 (01)09357123000037(11)230123(21)00348 350 (01)09357123000037(11)230328(21)00350. 351 (01)09357123000037(11)230413(21)00351 361 (01)09357123000037(11)230503(21)00361

Root Cause

Device Design

Action Taken

On April 4, 2025, Micro-X Ltd. issued a "Urgent Medical Device Correction" Notification to via E-Mail affected consignees. Micro-X ask consignees to take the following actions: " Inspect your stock immediately and quarantine affected stock on hand to prevent further use. " Circulate this notice to all users of this device so they are aware of the issue. " If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. " Please retain this letter with your system(s) documentation. " Please complete and return the attached response form to the Micro-X LTD in Australia: support@micro-x.com and copy djerkovic@micro-x.com to reconcile this process no later than 30 days from receipt. Completing this form confirms receipt of this notice, understanding of the issue, and the necessary actions to be taken.

More recalls by Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia

View all recalls by this company →