RecallDepth

thermage THERMAGE CPT SYSTEM REF TG-2B *** Solta Medical, Inc. thermage RF GENERATOR REF TG-2B *** Solta Medical, Inc. Dermatologic and general surgical procedures for electrocoagulation and hemostasis; " Non-invasive treatment of periorbital wrinkles and rhytids including upper and lower eyelids; " Non-invasive treatment of wrinkles and rhytids. The Thermage CPT System ( System ) consists of the RF Generator/ Cooling System, 16.00 cm2 Handpiece ( Body Handpiece ), Standard Handpiece with Vibration ( Face Handpiece ), optional electric Footswitch, accessory cables, Return Pad, Return Pad Cable, and Treatment Tips.

Company
Solta Medical Inc
Recall Initiated
February 16, 2023
Posted
April 7, 2023
Recall Number
Z-1354-2023
Quantity
1 unit
Firm Location
Bothell, WA

Reason for Recall

A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station.

Distribution

U.S. Distribution to state of: MD

Lot / Code Info

Model Number: TG-2B UDI-DI Code: 00850608002124 Serial Number: 001933

Root Cause

Process control

Action Taken

On 02/16/2023, the firm contacted customer via telephone e and then coordinate shipping out a loaner device as well as getting the unit (s/n: 001933) back for retest and possible recalibration. Documentation for the recall will be mailed with the loaner system to the customer. If you have any questions, contact Senior Director, Global Product Support at 425-420-2352 or email: Jody.Johnson@solta.com.