VASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).

This recall has been terminated (originally issued June 1, 2017).

Company
Solta Medical Inc
Recall Initiated
June 1, 2017
Terminated
November 30, 2020
Recall Number
Z-3048-2017
Quantity
7 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.

Distribution

US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.

Lot / Code Info

Reference No. 110-0032; Shipping P/N P007014-03; Device P/N P007023-03; Serial No. 001CW9, 000V1W, 000VBK, VAS10-0135-SB, 001FM0, 000TJ5, 00190Q.

Root Cause

Equipment maintenance

Action Taken

Consignees were telephoned on June 1, 2017, as initial recall notification. At that time, all consignees agreed to discontinue use of the affected product until it was returned or repaired. An Urgent Medical Device Recall letter dated June 14, 2017, requested that consignees that have not been contacted regarding this recall and have a device repaired between the dates of 09/08/2016 and 05/23/2017 contact Solta Medical immediately at 877-782-2286. Any questions regarding the recall can be directed to Solta Medical Product Support Team at 877-782-2286 or 510-259-5299, option 2.