Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.

Company
Maquet Cardiovascular, Llc
Recall Initiated
February 9, 2022
Recall Number
Z-0887-2022
Quantity
60
Firm Location
Wayne, NJ

Reason for Recall

Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.

Distribution

Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.

Lot / Code Info

UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.

Root Cause

Mixed-up of materials/components

Action Taken

On February 14, 2022, a customer notification letter was issued via FedEx. Customers are instructed to quarantine and return all affected product and notify all users within the hospital/facility. Please note affected lot numbers appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number, which is etched onto all 3 device components on side of handle and on underside above each pin. Customers are to complete and return the response form by email or fax. Distributors are to forward this document to customers for appropriate action. If you have any questions, please contact your Maquet Cardiovascular, LLC /Getinge representative or call the Getinge Customer Service at (888) 880-2874, Monday through Friday, between the hours of 6:00a.m. and 5:00 p.m. (Pacific Standard Time)