Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564

This recall has been terminated (originally issued October 20, 2021).

Company
Biomeme, Inc.
Recall Initiated
October 20, 2021
Terminated
March 5, 2024
Recall Number
Z-0452-2022
Quantity
11,607
Firm Location
Philadelphia, PA
Official Source
View on FDA website ↗

Reason for Recall

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).

Distribution

Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.

Lot / Code Info

All lots. Last lot of Kit distributed with the old IFU is 20210914.

Root Cause

Incorrect or no expiration date

Action Taken

On October 20, 2021, the firm sent an informal notice to customers titled "Product Update: Storing your RNA Process Control (RPC)". On November 4, 2021, the firm followed up with a formal notice titled "Medical Device Correction". The letter notified customers of the need to shorten the life of the RPC Buffer once resuspended from one week to TWO DAYS while stored at room temperature (15 degrees Celsius - 30 degrees Celsius). Actions to be taken by the Customer: Effective immediately, users of resuspended RPC Buffer should discard after two days when stored at room temperature. For longer-term storage, the RPC Buffer can be frozen once, at a temperature of -20C, for up to 3 months, and should not be frozen and thawed more than once. The expiration date of unopened RPC Kits will remain the same. Customers were also instructed to confirm receipt of the notification through a link provided by the firm. Customers should discard any other documents they may have on-hand that specifies the longer shelf-life stated in the IFU. When available, Biomeme will deliver updated versions of the IFU to all users by email.