Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a "baseboard" unit with firmware and associated software component that control motorized movements of the optical filter carriage used for PCR detection.

This recall has been terminated (originally issued January 5, 2021).

Company
Biomeme, Inc.
Recall Initiated
January 5, 2021
Terminated
June 11, 2021
Recall Number
Z-0940-2021
Quantity
147 systems
Firm Location
Philadelphia, PA
Official Source
View on FDA website ↗

Reason for Recall

Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.

Distribution

Domestic: FL, TX, IN, KY, OH, NY, CA, PA, MD. Foreign: Canada, Zimbabwe.

Lot / Code Info

Device Reference/Catalog Number (Black) - 1000003, (white) - 1000018. Software version 6.4.3  EUA - Emergency Use Authorization

Root Cause

Software design

Action Taken

On 01/05/2021 Biomeme, Inc., issued an Urgent Device Recall Correction for the latest build of the Franklin Real-Time PCR Thermocycler, which affects some devices built since September 21st, 2020. This software correction is to fix an issue where the device is unable to get past the beginning stages of the test.