HALYARD STERLING ZERO NITRILE POWDER-FREE EXAM GLOVES, Sizes XS, S, M, L, and XL
- Company
- O & M Halyard Inc
- Recall Initiated
- September 27, 2021
- Recall Number
- Z-0317-2022
- Quantity
- 2,928 cases (5,856,000 gloves)
- Firm Location
- Alpharetta, GA
Reason for Recall
All lots of the glove may discolor and lose elasticity upon aging, resulting in the inability of the user to don the gloves without tearing.
Distribution
Worldwide distribution - US Nationwide and the countries of EU/Canada.
Lot / Code Info
All lots of the following sizes: a. XS, REF 47459, UDI (01)20680651474593 b. S, REF 47460, UDI (01)20680651474609 c. M, REF 47461, UDI (01)20680651474616 d. L, REF 47462, UDI (01)20680651474623 e. XL, REF 47463, UDI (01)20680651474630
Root Cause
Component design/selection
Action Taken
O&M Halyard Inc has issued a voluntary recall of the Sterling Zero Nitrile Powder-free Exam Glove on 10/08/2021 by email or Fed Ex. The letter requested that the consignee segregate and destroy all affected product. Distributors were directed to notify their customers.