Orthopedic Pack, Model: 88041-12 AC2424601B 09/02/2029 Universal Pack, Model: 77165-06 AC2424902B 09/05/2029

Company
O & M Halyard Inc
Recall Initiated
July 17, 2025
Posted
September 4, 2025
Recall Number
Z-2520-2025
Quantity
1316
Firm Location
Alpharetta, GA

Reason for Recall

Surgical drape packs may have open seals, which may compromise the sterility of the product.

Distribution

Worldwide - US Nationwide distribution in the states of GA, AL, PA and the countries of MX.

Lot / Code Info

Model/UDI-DI/Lot(expiration): 88041-12/30680651880414/AC2424601B9(09/02/2029), 77165-06/30680651771651/AC2424902B(09/05/2029)

Root Cause

Process control

Action Taken

On 7/17/2025, Field Action Customer Notifications were emailed to customers who were asked to do the following: 1) DESTROY or DISCARD the product available in your inventory per the facility's standard disposal procedure for surgical packs. 2) COMMUNICATE this Field Action to staff to ensure awareness of it. 3) If any of the affected lot identified above was further distributed, please ensure all end-users are appropriately notified of this Field Action and maintain records of effectiveness. 4) Any adverse events associated with the use of the affected product should be reported to complaints@owens-minor.com. 5) RETURN the completed Field Action Response Form to GM-OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. 6) If credit is requested, please get contact with the Customer Service team and send email to GM-OMRA_RECALLS@owens-minor.com. If you have any questions, please email GM-OMRA_RECALLS@owens-minor.com using the Event # FA-2025-015 in the header of the response

More recalls by O & M Halyard Inc

View all recalls by this company →