smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, R, 5H, 115MM, REF 72573202
- Company
- Smith & Nephew Inc
- Recall Initiated
- November 1, 2021
- Recall Number
- Z-0366-2022
- Quantity
- 41 units
- Firm Location
- Memphis, TN
Reason for Recall
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
Distribution
US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
Lot / Code Info
Batch Number: 20HM10258, 20HM10259, 20HM10261, 20HM22791, UDI
Root Cause
Under Investigation by firm
Action Taken
The firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.