RecallDepth

META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690

Company
Smith & Nephew Inc
Recall Initiated
March 18, 2025
Posted
May 19, 2025
Recall Number
Z-1806-2025
Quantity
75
Firm Location
Memphis, TN

Reason for Recall

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

Distribution

Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.

Lot / Code Info

UDI-DI: 00885556132975. Lot: 24CT81426. Expiration: 25-Mar-2034

Root Cause

Process control

Action Taken

On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to FieldActions@smith-nephew.com If you have any questions email the firm at the email address above.