LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837

Company
Covidien Llc
Recall Initiated
December 6, 2021
Posted
January 13, 2022
Terminated
May 11, 2023
Recall Number
Z-0495-2022
Quantity
504 devices
Firm Location
Mansfield, MA

Reason for Recall

Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.

Distribution

Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.

Lot / Code Info

UDI: 20884521705880 10884521705883; Lot Number: 11320187X

Root Cause

Device Design

Action Taken

URGENT MEDICAL DEVICE RECALL notification letters dated 12/2021 were sent to customers. Customer Action Required: 1. Please immediately quarantine and discontinue use of the affected item code with the associated lot number listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated lot number must be returned. 3. If you have distributed the product item code and lot number listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory and email it to rs.anzrecalls@medtronic.com We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please do not hesitate to contact Azi Hosseini on +61 428 464 838, or azi.hosseini@medtronic.com