Puritan Bennett 560 Ventilator, PB560, CFN 4096600
- Company
- Covidien Llc
- Recall Initiated
- August 19, 2022
- Posted
- September 22, 2022
- Recall Number
- Z-1775-2022
- Quantity
- 5 US
- Firm Location
- Mansfield, MA
Reason for Recall
Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.
Distribution
Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.
Lot / Code Info
UDI-DI: 10884521196377 U.S. Serial Numbers: 40966N7229 40966N7231 40966N7334 40966N7335 40966N7340
Root Cause
Process control
Action Taken
Consignees were notified by letter delivered beginning August 19, 2022. Medtronic is advising that continued use of PB560 ventilators is appropriate when the actions listed in the recall letter are taken into consideration, unless this continued use is inconsistent with facility internal policies and procedures. All affected units require turbine replacement. Customers should contact Medtronic Service at 1-800-255-6774 to arrange for required service to replace the affected ventilator turbine component. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.