SDMS (SenTec Digital Monitoring System)

This recall has been terminated (originally issued October 17, 2016).

Company
Sentec Ag Ringstrasse 39 Therwil Switzerland
Recall Initiated
September 15, 2016
Posted
October 17, 2016
Terminated
June 23, 2017
Recall Number
Z-0088-2017
Quantity
75 (54 US; 21 =OUS)
Official Source
View on FDA website ↗

Reason for Recall

Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).

Distribution

Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.

Lot / Code Info

SN 306497 to SN 306571.

Root Cause

Component change control

Action Taken

Distributors were sent via e-mail a SenTec 'Mandatory initiated field correction action" letter dated September 15, 2016 . The letter provided a "Description of the issue, Affected products, Identify and localize affected products, Removal from use, Replacement of affected products at end customer & distributor / sub-distributor level". It also provided "action information for sub-distributors and affiliates, and end customers". For questions contact SenTec Inc. by e-mail (customersevice.usa@sentec.com) or phone (636-343-0000).