V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).

Company
Sentec Ag Ringstrasse 39 Therwil Switzerland
Recall Initiated
May 2, 2023
Posted
June 20, 2023
Recall Number
Z-1980-2023
Quantity
11 in the U.S.

Reason for Recall

The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.

Distribution

US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.

Lot / Code Info

Serial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.

Root Cause

Device Design

Action Taken

The recalling firm issued a recall letter dated 5/2/2023 via email on 5/2/2023. The letter explained the reason for recall also stating that the sensors pose no risk to patients but have a high chance of not functioning upon being connected to an SDM monitor. The recalling firm informed the consignee they want to replace the defective sensors. An RMA was enclosed to return the sensor and the shipping confirmation of its replacement.