PREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal Antigen enzyme immunoassay (EIA) is a screening or a semi-quantitative test system for the detection of capsular polysaccharide antigens of Cryptococcus neoformans in serum and cerebrosoinal fluid (CSF).

Company
Meridian Bioscience Inc
Recall Initiated
December 17, 2018
Terminated
April 30, 2020
Recall Number
Z-0732-2019
Quantity
184
Firm Location
Cincinnati, OH

Reason for Recall

An enzyme reagent included in the kit is not maintaining stability through claimed product expiration, resulting in the potential for false negative results when tested with patient specimens and the positive control.

Distribution

Worldwide Distribution - US Nationwide in the states of: AK, CA, CO, DE, IL, KS, KY, LA, MD, MO, NC, ND, NJ, NY, SC, TX, and VA. The products were distributed to the following foreign countries: India and Italy.

Lot / Code Info

Lot Numbers: 602096K089, 602096K090, 602096K091

Root Cause

Unknown/Undetermined by firm

Action Taken

Urgent: Voluntary Medical Device Recall notification letters dated 12/17/18 were distributed to customers. The letter identified the affected product, problem and actions to be taken. The letters instructed customers to review inventory for affected lots, forward the letter to any customers, contact the firm replacement product, complete the response form. For questions contact Meridian's Technical Service Department by telephone at 800-343-3858 or 513-271-3700 or by facsimile 513-272-5432.