RecallDepth

Durex Real Feel, synthetic polisoprene male condom, SKU 02340-89456-14 02340-89455-00 02340-89456-14 02340-89456-14 02340-89455-00 02340-95767-00 02340-89456-14 02340-89457-00 02340-89456-14 02340-89456-14 02340-89455-00 02340-89455-00 02340-95767-00 02340-95767-00

Company
Reckitt Benckiser Llc
Recall Initiated
December 19, 2018
Posted
March 9, 2019
Terminated
March 26, 2021
Recall Number
Z-0993-2019
Quantity
1761044
Firm Location
Parsippany, NJ

Reason for Recall

Five batches of the product have failed the specification of 1.0kPa for Burst Pressure.

Distribution

Nationwide domestic distribution. Foreign distribution to Barbados, Jamaica, Aruba, Guyana, Haiti, and St Maarten.

Lot / Code Info

1000314778 1000289702 1000322841 1000329839 1000317735 1000323210 1000318792 1000170515 1000240186 1000378288 1000183941 1000350794 1000270527 1000277098

Root Cause

Under Investigation by firm

Action Taken

On December 19, 2018, the firm issued Urgent Medical Device Recall letters to wholesale customers. The letter advised customers of the product issue, and reminded customers that the product labeling states, "Should a condom break or leak during use, seek medical assistance as soon as possible, at least within 72 hrs." Customers are asked to do the following: 1. Examine inventory in your distribution center immediately to determine if you have the specified batch in your warehouse inventory. 2. Segregate and return the specified batch. 3. A pre-paid return label is provided by the firm for your convenience to return the product. Contact Stericycle Expert Solutions at 888-843-0252 for assistance if you need additional labels or help with coordination of pickup and removal of product from your facility. Contact Information: Phone: 888-843-0252 Fax: 855-880-6687 Email: rbhealth3552@stericycle.com 4. Alternatively, you may arrange shipment of product directly to Stericycle at the following address: Attn: Event 3552 2670 Executive Drive, Suite A Indianapolis, IN 46241 Dock Hours are 8AM - 4:30PM Monday - Friday Please call to schedule a dock appointment 5. Finally, notify Stericycle representative by completing and returning the Business Response Form via email rbhealth3552@stericycle.com that you have: a. Checked your inventory in your distribution centers and confirm you have NO product of the specified batch. b. You have collected all units in your inventory and product has been shipped back in accordance with the instructions from your Stericycle representative. Please direct any questions or concerns with this recall to Stericycle Representative at 877-448-5308 and reference event 3552. For general inquiry not related to the recall, please contact RB Custo er Service Director, Robert Drennan at 201-220-8531.

More recalls by Reckitt Benckiser Llc

View all recalls by this company →