RecallDepth

Polyisoprene condomns Product Usage: The Durex Synthetic polyisoprene male condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted disease).

Company
Reckitt Benckiser Llc
Recall Initiated
August 6, 2018
Terminated
November 12, 2019
Recall Number
Z-0091-2019
Quantity
34 cartons/12 condoms each (408 condoms total)
Firm Location
Parsippany, NJ

Reason for Recall

Product marketed in U.S. does not have sufficient data to ensure that it will meet the internal burst pressure specifications at the end of 5 yrs shelf-life

Distribution

US in the states of NY

Lot / Code Info

Lot Number - 1000490884

Root Cause

Nonconforming Material/Component

Action Taken

On August 6,. 2018, Reckitt Benckiser issued Urgent Medical Device Recall notices to customers. Customers were advised to take the following action: Wholesale Customers: 1. Examine inventory and quarantine affected product 2. Contact customers and notify them of the recall and provide them with the recall instructions (copy of the letter). 3. Customers are encouraged to contact Stericycle Expert Solutions at 877-448-5308 for assistance if additional return labels are required or help with coordination of pickup and removal of product from your facility.