PREMIER EHEC and PREMIER EHEC Bulk.
- Company
- Meridian Bioscience Inc
- Recall Initiated
- November 8, 2016
- Posted
- January 17, 2017
- Terminated
- August 16, 2017
- Recall Number
- Z-1033-2017
- Quantity
- 4,747 EHEC + 180 EHEC Bulk
- Firm Location
- Cincinnati, OH
Reason for Recall
Internal testing has demonstrated that the indicated kit lots produce false positive results when tested with some broth specimens.
Distribution
Product was shipped to the following states: AL. AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WY. Product was also sent to the following countries: CANADA, GERMANY, HONG KONG, ITALY, JAPAN & NEW ZEALAND.
Lot / Code Info
PREMIER EHEC Product Code 608096: Lot # 608096H219, Exp Date 9/7/2017, Detection Antibody Reagent, Lot # 8603.138 Exp Date 9/30/2017; Lot # 608096H222, Exp Date 12/20/2017, Detection Antibody Reagent, Lot # 8603.14 Exp Date 1/12/2018; Lot # 608096H224, Exp Date 9/7/2017, Detection Antibody Reagent, Lot # 8603.141 Exp Date 1/29/2018; Lot # 608096H225, Exp Date 12/20/2017, Detection Antibody Reagent, Lot # 8603.142, Exp Date 2/5/2018 & Lot # 608096H226, Exp Date 3/6/2018, Detection Antibody Reagent, Lot # 8603.143, Exp Date 3/22/2018. PREMIER EHEC Bulk Product Code 608096BU: Lot # 608096H220 Exp Date 9/7/2017, Detection Antibody Reagent, Lot # 8603.138, Exp Date 9/30/2017; Lot #, 608096H223, Exp Date 12/20/2017, Detection Antibody Reagent, Lot # 8603.14, Exp Date 1/12/2018 & Lot # 608096H227, Exp Date 3/6/2018, Detection Antibody Reagent, Lot # 8603.143, Exp Date 3/22/2018.
Root Cause
Under Investigation by firm
Action Taken
On November 7, 2016 the firm sent Voluntary Medical Device Recall Letters and reply forms to their customers.