Panther Fusion Tube Tray, Part: FAB-15004, REF: PRD-04000

Company
Hologic, Inc
Recall Initiated
August 4, 2020
Terminated
December 23, 2024
Recall Number
Z-2990-2020
Quantity
607
Firm Location
San Diego, CA

Reason for Recall

Tube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay.

Distribution

U.S. :ID, SC, VT, CA, IL, UT, NJ, MA, TX, AL, CO, FL, ME, NY, VA, WA, WI, OR, DC, GA, NC, HI, CT, OK, TN, O.U.S.: CA, SE, GB, DE, DO, IT, ES, NO, BE, NL, FR, DK, AT

Lot / Code Info

Box/ Tray lots: 271436/ 618539,  271438/ 620899,  272089/ 620899,  272959/ 621837,  272970/ 621256,  272971/ 621837,  274670/ 625351

Root Cause

Under Investigation by firm

Action Taken

On 08/04/20, Medical Device Recall Notifications were mailed to Laboratory Managers, Site Administrators, and Operators. Customers were asked to do the following: 1. Discontinue use. 2. Immediately check inventory and segregate affected devices. 3. Complete the customer response form. Document the remaining inventory. Destroy the segregated inventory. Return the customer response form even if there is no affected inventory present. 4. There is no other specific actions to the assays. 5. Recalling firm will ship replacement product based on the customer response form information.